Product Name |
Rat Testo(Testosterone) ELISA Kit |
Catalogue No. |
4570ELK |
Alternative Names: |
T; Testesteron; Testeronun |
Assay Type |
Competitive Inhibition |
Sensitivity |
6.49 pg/mL |
Range |
156.25-10000 pg/mL |
Sample Type |
serum, plasma, other biological fluids |
Assay Length |
2h |
Research Area |
Endocrinology;Reproductive science;Hormone metabolism; |
Test principle |
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Testosterone(Testosterone) protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Testosterone(Testosterone). Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Testosterone(Testosterone) in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard Curve |
Concentration (ng/mL) |
OD |
Corrected OD |
10000.00 |
0.189 |
— |
5000.00 |
0.427 |
— |
2500.00 |
0.527 |
— |
1250.00 |
0.913 |
— |
625.00 |
1.197 |
— |
312.50 |
1.583 |
— |
156.25 |
1.813 |
— |
0.00 |
2.226 |
— |
|
Precision |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. |
Recovery |
Matrices listed below were spiked with certain level of recombinant Testo and the recovery rates were calculated by comparing the measured value to the expected amount of Testo in samples.
Matrix |
Recovery range |
Average |
serum(n=5) |
82-95% |
88% |
EDTA plasma(n=5) |
80-95% |
87% |
Heparin plasma(n=5) |
86-99% |
89% |
|
Linearity |
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Testo and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix |
1:2 |
1:4 |
1:8 |
1:16 |
serum(n=5) |
80-101% |
87-96% |
86-97% |
88-95% |
EDTA plasma(n=5) |
79-93% |
89-97% |
85-94% |
87-101% |
Heparin plasma(n=5) |
85-94% |
87-96% |
89-101% |
82-90% |
|
Note |
For Research Use Only |
Reviews
There are no reviews yet.