Osteocalcin (OC) Elisa Kit

Catalog Number BSEK-076C
Sensitivity 0.249ng/mL
Range 0.625-40ng/mL
Species Canis
Sample Type Serum, plasma, tissue homogenates and other biological fluids
Test Method Double-antibody Sandwich
Assay Length 3h
Size 96 T | 48 T
Price POR
Contact Us
SKU: BSEK-076C Categories: ,

Specificity

  • This assay has high sensitivity and excellent specificity for detection of Osteocalcin (OC).
  • No significant cross-reactivity or interference between Osteocalcin (OC) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Osteocalcin (OC) and the recovery rates were calculated by comparing the measured value to the expected amount of Osteocalcin (OC) in samples.

Matrix Recovery range Average
Serum (n=8) 95-105% 99%
EDTA plasma (n=8) 79-97% 90%
Heparin plasma(n=5) 83-99% 93%

Precision

  • Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Osteocalcin (OC) were tested 20 times on one plate, respectively.
  • Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Osteocalcin (OC) were tested on 3 different plates, 8 replicates in each plate.
  • CV(%) = SD/meanX100
  • Intra-Assay: CV<10%
  • Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Osteocalcin (OC) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 95-105% 96-103% 79-96% 94-102%
EDTA plasma(n=5) 78-96% 96-103% 91-103% 87-99%
Heparin plasma(n=5) 89-102% 78-88% 93-101% 79-93%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Test Principle

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Osteocalcin (OC). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Osteocalcin (OC). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Osteocalcin (OC), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Osteocalcin (OC) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

 