Human Testo(Testosterone) ELISA Kit

Catalogue No ELK8525
Standard 4000 pg/mL
Range 62.5-4000 pg/mL
Sensitivity 17.7 pg/mL
Reactivity Human
Contact Us
SKU: ELK8525 Categories: ,
Product Name Human Testo(Testosterone) ELISA Kit
Alias T; Testesteron; Testeronun
Assay Type Competitive Inhibition
Sample Type serum, plasma, other biological fluids
Assay Length 2h
Research Area Endocrinology; Reproductive science; Hormone metabolism

Standard Curve

Concentration (ng/mL) OD-1 Corrected OD
4000.00 0.217 -
2000.00 0.385 -
1000.00 0.672 -
500.00 0.878 -
250.00 1.152 -
125.00 1.596 -
62.50 1.624 -
0.00 2.256 -

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Testo and the recovery rates were calculated by comparing the measured value to the expected amount of Testo in samples.

Matrix Recovery Range Average Recovery
serum(n=5) 78-95% 86%
EDTA plasma(n=5) 80-96% 88%
Heparin plasma(n=5) 80-92% 85%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Testo and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-101% 87-98% 85-94% 87-101%
EDTA plasma(n=5) 86-93% 91-105% 87-99% 99-105%
Heparin plasma(n=5) 92-101% 89-97% 85-94% 87-101%

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Testosterone(Testo) protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Testosterone(Testo). Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Testosterone(Testo) in the samples is then determined by comparing the OD of the samples to the standard curve.