🚀 Multiplex Customized ELISA Kits – Accuracy, Efficiency, Innovation in Every Assay!

Human IgE (Immunoglobulin E) Elisa Kit

Human IgE (Immunoglobulin E) Elisa Kit

Catalogue No BSEK-261H
Species Human
Size 96 T | 48 T
Price POR
Contact Us
SKU: BSEK-261H Categories: ,
Human IgE (Immunoglobulin E) ELISA Kit
Sensitivity 6.34 pg/mL
Range 78.13-5000pg/mL
Uniprot ID P01854
Alternative Names IgE; immunoglobulin heavy constant epsilon
Detection Method Sandwich
Standard 5000pg/mL
Sample Type Serum, Plasma, Tissue homogenate and Other biological samples
Sample Volume 100μL
Reaction Time 3.5H
Research Area Signal transduction; Hematology; Cardiovascular biology; Neuro science
Principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IgE. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IgE and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IgE, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human IgE. You can calculate the concentration of Human IgE in the samples by comparing the OD of the samples to the standard curve.
Human IgE (Immunoglobulin E) ELISA Kit
Sensitivity 1.5ng/ml
Range 6ng/ml-400ng/ml
Detection Method Quantitative Detection
Wavelength 450nm
Standard 540ng/ml
Sample Volume 10ul
Sample Type use with serum, plasma, culture media, or any biological fluids
Principle The quantitative detection of Human IgE antigen is based on the ELISA (Enzyme-Linked Immunosorbent Assay) technique. The microplate wells are pre-coated with an antibody specific to Human IgE. After incubation, the corresponding antigen in the sample or standards will bind to the immobilized antibody. After washing the microplate wells to remove all unbound sample material, horseradish peroxidase (HRP) labeled antibody conjugate is added, which binds to the antigen, forming antibody-antigen-enzyme-labeled antibody complex upon incubation. In a second wash step, all unbound material is removed. The immune complex formed by the bound conjugate is visualized by adding Tetramethylbenzidine (TMB) substrate, which gives a blue color reaction in wells containing the complex. A stop solution containing acid terminates the reaction, producing a yellow end-point coloration proportional to the amount of specific antigen present in the sample. Absorbance is measured at 450 nm using an ELISA microwell plate reader. Analysis is completed by comparing the OD values of samples to the standard curve

Intra-assay Precision (Precision within an assay): 3 samples with low, middle, and high levels of Human IgE were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, middle, and high-level Human IgE were tested on 3 different plates, 8 replicates in each plate.

CV(%) = SD/mean X 100
Intra-Assay: CV<10%
Inter-Assay: CV<12%