Bovine Catalase (CAT) Elisa Kit

Catalogue No BSEK-095B
Species Bovine
Sensitivity 0.01ng/ml
Range 0.05ng/ml-4ng/ml
Size 96 T | 48 T
Price POR
Contact Us
SKU: BSEK-095B Categories: ,

Purpose

Our Bovine Catalase (CAT) ELISA kit is to assay CAT levels in Bovine serum, plasma, culture media or any biological fluid. Enzyme immunoassay for the quantitative detection of Bovine CAT antigen in serum, plasma, culture media, or any biological fluid.

Bovine Catalase (CAT) Elisa Kit - Details
Wavelength 450nm
Standard 5.4ng/ml
Sample Volume 10ul
Samples serum, plasma, culture media, or any biological fluids
Detection Method Sandwich
Storage 2-8℃
Principle
The quantitative detection of Bovine CAT antigen is based on the ELISA (Enzyme-Linked Immunosorbent Assay) technique. The microplate wells are pre-coated with an antibody specific to Bovine CAT. The corresponding antigen in the sample or standards binds the immobilized antibody after incubation. After washing the microplate wells to remove all unbound sample material, horseradish peroxidase (HRP) labeled antibody conjugate specific to CAT is added, whichbinds the captured antigen, forming antibody-antigen-enzyme-labeled antibody complex upon incubation. In a second wash step, the unbound conjugate is removed. The immune complex formed by the bound conjugate is visualized by adding Tetramethylbenzidine (TMB) substrate, which gives a blue color reaction in wells containing the complex. A stop solution containing acid terminates the reaction, producing a yellow end-point coloration proportional to the amount of specific antigen present in the sample. Absorbance is measured at 450 nm using an ELISA microwell plate reader. Analysis is completed by comparing the OD values of samples to the standard curve.

Precision

  • Intra-assay Precision (Precision within an assay): 3 samples with low, middle, and high levels of Bovine CAT were tested 20 times on one plate, respectively.
  • Inter-assay Precision (Precision between assays): 3 samples with low, middle, and high-level Bovine CAT were tested on 3 different plates, 8 replicates in each plate.
  • CV(%) = SD/mean X 100
  • Intra-Assay: CV<10%
  • Inter-Assay: CV<12%