Test Principle
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with 11-KT. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to 11-KT. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of 11-KT in the samples is then determined by comparing the OD of the samples to the standard curve.
Product Name |
11-KT(11-keto Testosterone) ELISA Kit |
Alternative Names |
11-keto Testosterone |
Assay Type |
Competitive Inhibition |
Standard |
10000 pg/mL |
Sample Type |
serum, plasma and other biological fluids |
Assay Length |
2h |
Research Area |
Endocrinology, Reproductive Science, Hormone Metabolism |
Standard Curve
Concentration (pg/mL) |
OD |
Corrected OD |
10000.00 |
0.191 |
- |
5000.00 |
0.397 |
- |
2500.00 |
0.715 |
- |
1250.00 |
0.823 |
- |
625.00 |
1.257 |
- |
312.50 |
1.572 |
- |
156.25 |
1.769 |
- |
0.00 |
2.232 |
- |
Precision
- Intra-assay Precision (Precision within an assay):CV%<8%
- Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
- Inter-assay Precision (Precision between assays):CV%<10%
- Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant 11-KT and the recovery rates were calculated by comparing the measured value to the expected amount of 11-KT in samples.
Matrix |
Recovery range |
Average |
serum(n=5) |
90-105% |
97% |
EDTA plasma(n=5) |
86-99% |
92% |
Heparin plasma(n=5) |
80-95% |
86% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 11-KT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix |
1:2 |
1:4 |
1:8 |
1:16 |
serum(n=5) |
79-94% |
87-96% |
87-101% |
82-90% |
EDTA plasma(n=5) |
87-95% |
78-91% |
97-103% |
95-101% |
Heparin plasma(n=5) |
85-93% |
88-95% |
96-102% |
83-97% |
Reviews