• Get In Touch (814) 262-7331
  • 633, Napoleon Street Johnstown, Pennsylvania PA,15901

Alternative Names:  3-Nitro-L-Tyrosine; 3-Nitrotyrosine; Nitrotyrosine
Catalogue No. 6687ELK
Size 96T
Reactivity General
Range 4.69-300 ng/mL
Sensitivity 1.35 ng/mL

Description

Alternative Names:  3-Nitro-L-Tyrosine; 3-Nitrotyrosine; Nitrotyrosine
Catalogue No. 6687ELK
Size 96T
Reactivity General
Range 4.69-300 ng/mL
Sensitivity 1.35 ng/mL
Assay Type Competitive Inhibition
Sample Type serum, plasma and other biological fluids
Assay Length 2h
Research Area Metabolic pathway; Infection immunity; Rheumatology;
Test principle This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Nitrotyrosine(NT) protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Nitrotyrosine(NT). Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Nitrotyrosine(NT) in the samples is then determined by comparing the OD of the samples to the standard curve.
Standard Curve

Concentration (ng/mL) OD Corrected OD
300.00 0.217
150.00 0.467
75.00 0.793
37.50 0.967
18.75 1.236
9.38 1.596
4.69 1.861
0.00 2.265
Precision Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery Matrices listed below were spiked with certain level of recombinant 3-NT and the recovery rates were calculated by comparing the measured value to the expected amount of 3-NT in samples.

Matrix Recovery range Average
serum(n=5) 92-107% 99%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 78-92% 84%
Linearity The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 3-NT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-93% 91-105% 88-104% 97-103%
EDTA plasma(n=5) 85-95% 79-96% 89-101% 85-97%
Heparin plasma(n=5) 87-96% 81-93% 86-98% 91-103%