• Get In Touch (814) 262-7331
  • 633, Napoleon Street Johnstown, Pennsylvania PA,15901

Product name: Human TJP1(Tight Junction Protein 1) ELISA Kit
Catalogue No.: ELK3029
Alternative Names: ZO1; Zona Occludens 1; Zona occludens protein 1; Zonula occludens protein 1
Assay Type: Sandwich
Sensitivity: 0.059 ng/mL
Standard: 10 ng/mL
Range: 0.16-10 ng/mL

Description

Product name: Human TJP1(Tight Junction Protein 1) ELISA Kit
Reactivity: Human
Alternative Names: ZO1; Zona Occludens 1; Zona occludens protein 1; Zonula occludens protein 1
Assay Type: Sandwich
Sensitivity: 0.059 ng/mL
Standard: 10 ng/mL
Range: 0.16-10 ng/mL
Sample Type: tissue homogenates, cell lysates and other biological fluids
Assay Length: 3.5h
Research Area: Signal transduction;Tumor immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Tight Junction Protein 1(TJP1). Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Tight Junction Protein 1(TJP1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Tight Junction Protein 1(TJP1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Tight Junction Protein 1(TJP1) in the samples is then determined by comparing the OD of the samples to the standard curve.
Standard curve

Concentration (ng/mL) OD Corrected OD
10.00 2.082 1.988
5.00 1.625 1.531
2.50 1.175 1.081
1.25 0.827 0.733
0.63 0.583 0.489
0.32 0.375 0.281
0.16 0.206 0.112
0.00 0.094 0.000
Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant TJP1 and the recovery rates were calculated by comparing the measured value to the expected amount of TJP1 in samples.
Matrix Recovery range Average
serum(n=5) 85-97% 91%
EDTA plasma(n=5) 92-105% 98%
Heparin plasma(n=5) 82-95% 88%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TJP1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-102% 86-92% 81-93% 89-104%
EDTA plasma(n=5) 87-99% 85-96% 87-101% 89-102%
Heparin plasma(n=5) 85-92% 80-96% 87-98% 95-102%
Note For Research Use Only

Reviews

There are no reviews yet.

Be the first to review “Human TJP1(Tight Junction Protein 1) ELISA Kit”

Your email address will not be published. Required fields are marked *